Skip to content

Search these docs, or ask Revo a question — answers link the pages they came from.

Hamilton Incubator Shaker

The Hamilton Incubator Shaker (HIS) incubates and shakes several plates at once, in independent slots. Hamilton states it pairs with their automated liquid handlers and that multiple HIS units may be added to increase throughput.

Temperature control4 °C above ambient to 60 °C
Shaking speedUp to 1200 rpm
Shaking orbit3 mm orbital
Plate heightUp to 48 mm
CapacityUp to four independent slots, by configuration
Revolution controlSet temperature, set shaking speed, set duration, start, read status

Hamilton publishes the following dimensions for the Incubator Shaker.

Height33.4 cm (1 ft, 1.15 in)
Width18 cm (7.09 in)
Depth27.4 cm (10.79 in)

Hamilton offers three configurations, chosen by labware type:

  • 4 lidded microplates
  • 2 lidded microplates + 1 lidded deep-well plate
  • 2 lidded deep-well plates
  • Fork mechanism — an integrated automated loading mechanism that lets CO-RE Paddles on the liquid handler load and unload plates without additional grippers
  • Real-time temperature logging, which Hamilton describes as ensuring consistent conditions throughout a run
  • Hamilton notes the unit minimises evaporation and shields samples from light

Hamilton lists genomics, nucleic acid extraction, PCR set-up, NGS library prep and single cell analysis as common applications.

  • Set temperature — set the target incubation temperature
  • Set speed — set the shaking speed
  • Set time — set the duration for the operation
  • Start — begin incubating and shaking with the values already set
  • Get status — read the unit’s current state, so a schedule can wait on it

Revolution talks to the HIS over a serial connection, and the device carries a configured plate type.

  • The unit must be cabled to the Revolution host and its COM port known before the device is initialised.
  • The configured plate type must match the labware and the unit’s configuration — a deep-well configuration and a microplate method are not interchangeable.
PropertyPurpose
COM portSerial port the unit is connected on
Plate typePlate type the device is configured for

If you need more of this instrument driven from a schedule, get in touch — the driver is extended on demand.