Search these docs, or ask Revo a question — answers link the pages they came from.
Cytomat 6000
The Cytomat 6000 is a mid-size random-access plate store built around a rotary carousel — the smallest of Revolution’s three rotary Cytomat stores by footprint, alongside the Cytomat 10C and Cytomat 24 (PH). Revolution’s driver for it does not expose any atmosphere or shaking control; only the Cytomat 2 does. Thermo’s own C600x-series and 6001 incubators at this size do offer temperature and humidity control, but that isn’t wired up as a schedulable operation on this device class.
At a glance
Section titled “At a glance”| Capacity | Up to 189 standard (96-/384-well) SBS microplates |
| Default grid (Revolution simulation layout) | 9 columns × 21 rows |
| Atmosphere / shaking control | Not exposed by this driver |
| Revolution control | The store’s turntable and shovel transfer mechanism; self-scan inventory |
Specifications
Section titled “Specifications”Thermo Fisher’s published figure for this size of Cytomat is up to 189 standard SBS microplates, consistent across the 6000/6001/C600x listings for this footprint. A breakdown by plate type (1536-well, deep-well, etc.) specific to this size could not be confidently sourced, so it isn’t given here.
Revolution’s default simulation layout — 9 columns × 21 rows, which multiplies out to 189 — lines up with that published figure. This is a default used for simulation and commissioning, not a guaranteed physical maximum; actual capacity depends on plate height and the shelf pitch fitted.
Supported methods
Section titled “Supported methods”Transfer mechanism
Section titled “Transfer mechanism”- Extend shovel / retract shovel
- Rotate shovel to carousel / rotate shovel to setdown — the shovel moves between two fixed positions on this device, not to an arbitrary location
- Rotate turntable to store location
- Move height motor to above / below store location
- Move to wait position
- Move transfer station to basic position / to position 2 — only meaningful where a
transfer station is fitted (
HasTransferStationdevice property) - Open / close automatic lift door
- Initialise hardware
This is the store’s own mechanism exposed step by step. A schedule should normally rely on the scheduler’s standard pick/place mechanism to store and retrieve plates and leave it at that — the individual movements above exist for commissioning, teaching and fault recovery, where you need to place the shovel or the turntable exactly rather than move a plate.
Inventory and status
Section titled “Inventory and status”- Start self scan / Start self scan (full)
- Show status / Show extended status
- Show extended options settings
A self-scan rebuilds the inventory from what is physically present. Run one after any manual loading, or Revolution’s view of the contents is only as good as its bookkeeping.
Integration
Section titled “Integration”Prerequisites
Section titled “Prerequisites”- The store must be reachable from the Revolution host.
- Plate type must suit the store’s shelf pitch; capacity depends on plate height.
- This driver does not read back or set temperature, humidity, CO₂ or O₂ — plan atmosphere- sensitive work around that rather than assuming set points exist to check.
- Run a self-scan after manual loading.
If you need more of this instrument driven from a schedule, get in touch — the driver is extended on demand.